8.3.10 Absorbance Cuvette (infinite
Accuracy
Measurement Parameters:
Material:
Calculation of Accuracy:
2008-07
®
For the
with the option 'Cuvette' the following test may be
infinite
M200
performed to prove the accuracy specification:
Perform the following measurement to determine the absorbance accuracy:
Parameter
Reading Mode
Measurement Wavelength
Number of flashes
Time between Move and Read
Cuvette Type
®
Starna
reference material RM-N1N35N + D3 cuvette
(for more information please refer to www.starna.co.uk)
Calculate the deviation of the measured value from the reference value supplied
with the calibrated cuvette.
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
®

M200 only)

Setting
Absorbance
1) 440 nm
2) 635 nm
25
0
Calibrated cuvette, e.g. Starna
RM-N1N35N + a D3 Cuvette
8. Quality Control
®