5.4.1

Absorbance Optics MTP

5.4.2

Absorbance Detection MTP

2008-07
A fiber bundle guides the light from the excitation monochromator system to the
absorbance MTP optics.
The absorbance optics consists of a pair of lenses focusing the light beam into
the well of the microplate.
The spot size of the absorbance light beam is 0.7 mm in diameter.
A silicon photodiode is used for the measurement of the transmitted light. It is
sensitive to a wide range of wavelengths. The photodiode is well suited for the
light levels being encountered with absorbance measurements up to 3 OD.
For absorbance measurement of nucleic acids of
small volumes (2µl) use Tecans's NanoQuant Plate.
With this device it is possible to measure
16 different samples within one measurement.
For further information please contact
your local Tecan distributor or visit:
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
Note
www.tecan.com
5. Optical System
77