2. General Description
2.3.6
Injector Reagent Compatibility
36
The injector system of the infinite
•
Teflon (PTFE): Tubing, valve plug, seal
•
KelF: Valve body
•
SC05: Injector needles
Please refer to the following list for reagent compatibility. The column 'Rating'
indicates the compatibility with the reagents listed in the 'Chemical' column.
Rating 'A' indicates a good compatibility with the injector system. Chemicals with
a rating 'D' must not be used with the infinite
damage the injector system.
Chemical
Acetic Acid < 60%
Acetonytrile
Butyl Amine
Carbon Tetrachloride (dry)
Chloroform
Diethyl Ether
Dimethyl Formamide
Ethanolamine
Ethanol
Ethylene Diamine
Furfural
Hexane
Hydrofluoric Acid
Methanol (Methyl Alcohol)
Monoethanolamine
Potassium Hydroxide (Caustic Potash)
Potassium Hypochlorite
Sodium Hydroxide
Sodium Hypochlorite
Concentrated Sulfuric Acid
Diluted Sulfuric Acid (Concentration ≤ 1 N)
Tetrahydrofuran
Water, Deionized
Water, Distilled
Water, Fresh
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
®
F/M200 consist of the following materials:
®
injectors. They will severely
Rating
A
A
D
D
A
D
A
D
A
D
D
A
D
A
D
D
D
D
D
D
A
A
A
A
A