Open Cover - Agilent Technologies G1314B User Manual

Agilent 1260 infinity variable wavelength detector
Table of Contents

Advertisement

Probable cause
1
2
3

Open Cover

Error ID: 0205
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If
the foam is removed during operation, the lamp and grating drive power is
switched off, and the error message is generated.
Probable cause
1
2
3
Agilent 1260 Infinity VWD User Manual
Fan cable disconnected.
Defective fan.
Defective main board.
The top foam was removed during
operation.
Foam not activating the sensor.
Defective sensor or main board.
Error Information
General Error Messages
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
7
93