Open Cover; Eri Messages - Agilent Technologies InfinityLab LC Series User Manual

Variable wavelength detectors
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

7
Error Information
General Error Messages

Open Cover

Error ID: 0205
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the foam
is removed, the fan is switched off, and the error message is generated.
Probable cause
1
2
3

ERI Messages

Error ID: 11120 (+5 V) , 11121 (+25 V)
The ERI (Enhanced Remote Interface) provides two error events related to over
current situations on the +5 V and +24 V lines.
Probable cause
1
Agilent InfinityLab LC Series VWD User Manual
The top foam was removed during operation. Please contact your Agilent service representa-
Foam not activating the sensor.
Defective sensor or main board.
The load on the ERI is too high.
Suggested actions
tive.
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Suggested actions
Reduce the load.