Open Cover; Cover Violation - Agilent Technologies 1260 User Manual

Infinity diode array and multiple wavelength detector
Hide thumbs Also See for 1260:
Table of Contents

Advertisement

2 If it is desired that the system has only one leak drain tube, then it's

Open Cover

The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the
foam is removed, the fan is switched off, and the error message is generated.
Probable cause
1
2
3

Cover Violation

The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the
foam is removed while the lamps are on (or if an attempt is made to switch on
for example the lamps with the foam removed), the lamps are switched off,
and the error message is generated.
Probable cause
1
2
Agilent 1260 Infinity DAD and MWD User Manual
possible in increase the length of the small plastic tube which is connected
to port# 4 of the sampler switching valve. This tube can then be taken to
waste separately. The tube which normally serves as the detector cell outlet
tube can be used for this purpose.
The top foam was removed during operation. Reinstall the top foam.
Foam not activating the sensor.
Dirty or defective sensor.
The top foam was removed during operation. Please contact your Agilent service
Foam not activating the sensor.
Error Information
General Error Messages
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
representative.
Please contact your Agilent service
representative.
8
149