Display Subsystem; Table 13-1 Scpi Commands - Display - Keithley 6485 Instruction Manual

Picoammeter
Hide thumbs Also See for 6485:
Table of Contents

Advertisement

13-2
DISPlay, FORMat, and SYSTem

DISPlay subsystem

The commands in this subsystem are used to control the display over the bus.
Table 13-1
SCPI commands — display
Command
:DISPlay
:DIGits <n>
:DIGits?
:ENABle <b>
:ENABle <b>
[:WINDow[1]]
:TEXT
[:DATA] <a>
[:DATA]?
:STATe <b>
:STATe?
Note: *RST and SYSTem:PRESet have no effect on the display circuitry and user-defined text messages.
A) DISPlay:ENABle <b>
With front panel circuitry turned off, the instrument operates at a higher speed. While dis-
abled, the display is frozen and all front panel controls (except LOCAL) are disabled. Nor-
mal display operations can be resumed by using :ENABle to enable the display, pressing
the LOCAL key, or cycling power.
B) DISPlay:TEXT[:DATA] <a>
Message Types:
The display message can be up to 12 characters (ASCII) long. A space is counted as a
character. Excess message characters result in an error. Note that for the string type, the
message must be enclosed by single or double quotes.
An indefinite block message must be the only command in the program message, or the
last command in the program message. If you include a command after an indefinite block
Description
Set display resolution; 4 to 7.
Query display resolution.
Turn front panel display on or off.
Query front panel display status.
Path to control user text messages:
Define ASCII message "a" (up to 12 characters).
Query defined ASCII message.
Enable or disable text message mode.
Query text message status.
String
Indefinite Block
Definite Block
where: Y = number of characters in message (up to 12)
Model 6485 Picoammeter Instruction Manual
'aa...a' or "aa...a"
#0aa...a
#XYaa...a
X = number of digits that make up Y (1 or 2)
Default
Ref
6
Section 4
(see Note)
A
(see Note)
B
C

Advertisement

Table of Contents
loading

Table of Contents