Table 18-5 Route Command Summary; Table 18-6 Sense Command Summary - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

18-10
SCPI Command Reference
Table 18-5
ROUTe command summary
Command
:ROUTe
:TERMinals <name>
Select in/out terminals: (FRONt or REAR).
:TERMinals?
Query in/out terminals.
Table 18-6
SENSe command summary
Command
[:SENSe[1]]
Sense 1 Subsystem:
:DATA
Path to SENSe[1] data.
[:LATest?]
Return only most recent reading.
:FUNCtion
Select measurement function(s):
:CONCurrent <b>
Enable or disable ability to measure more
:CONCurrent?
Query concurrent state.
[:ON] <function list>
Specify functions to enable
:ALL
:COUNt?
[:ON]?
Returns list of functions that are enabled.
:OFF <function list>
Specify functions to disable:
:ALL
:COUNt?
:OFF?
Returns list of functions that are disabled.
:STATe? <name>
Query state of specified function: 1 (on)
1. For the Model 2430 Pulse Mode, concurrent measurements are always disabled. This command is invalid (error +831).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2400 Series SourceMeter
Description
Description
than one function simultaneously. When
1
disabled, volts function is enabled.
(VOLTage[:DC], CURRent[:DC], or
RESistance).
Enable all functions (concurrent
enabled) or enable ohms function
1
(concurrent disabled).
Query number of functions that are
enabled.
(VOLTage[:DC], CURRent[:DC], or
RESistance).
Disable all measurement functions.
Query number of functions that are
disabled.
or 0 (off).
®
User's Manual
Default
Source
parameter SCPI
memory
FRONt
Default
Source
parameter SCPI
memory
ON
CURRent
VOLTage,
RESistance

Advertisement

Table of Contents
loading

Table of Contents