Instrument Teach Wizard - Clearance Position Window
Use the Clearance Position window to teach the Pick Clearance
and PlaceClearance positions for the instrument position.
The preferred method is to click the Move Up 1 CM button until the
bottom of the tallest material used in the instrument will clear the
plate locators in the instrument. This method is more accurate
because it ensures that the Clearance position is directly above the
Pick/Place position.
Alternatively, you can place the tallest material used in the system
into the position you just taught, and then place the material used
for position teaching on top of it.
Clicking Next creates the clearance positions on the Positions tab.
Grip buttons - Open and close the gripper if you neeed to grip or
release a plate.
Servo Control buttons - Enables or disables the robot axes.