Changing the Twister II Fuse(s) (Continued)
4
5
P/N 113564 Rev. 2
Remove the single 3Amp fuse or the two 1.6Amp fuses.
Figure 84
shows the fuse location for a single 3Amp fuse for
100/120V configuration.
two 1.6Amp fuses for 240V configuration.
Figure 84. 100/120V Fuse Configuration
Figure 85. 240V Fuse Configuration
To convert the Twister II fuse configuration for 240V~ operation:
a
Using the small, flat-blade screwdriver, loosen the fuse block
screw 2 or 3 turns, but do not remove it.
b Gently lift the fuse block away from the screw and pedestal
and then remove as shown in
Figure 86. Removing the Fuse Block
Twister II Robot User's Manual
Twister II - Maintenance 106
Figure 85
shows the fuse locations for
Fuse Location (3Amp)
Fuse Block
Screw
Fuse Locations (1.6Amp)
Figure
86.