Adjusting the Instrument Positions

If the robot does not put the materials in the correct position in an
instrument, the instrument or robot may have been moved slightly
out of position. This can sometimes be corrected by moving the
instrument back into position as described below.
1
2
3
If repositioning the instrument does not solve the problem, you
should re-teach the instrument position, as described on
For problems with rack positions, you should re-teach the rack
positions, as described on
If you continue to have problems, contact Caliper Technical Support
(see
P/N 113564 Rev. 2
Place ONE material in the instrument. If adjusting an instrument
with a drawer that opens and closes, ensure that the drawer is
fully open.
Open the robot software in standalone mode. Use the existing
positions to move the robot to the instrument position.
Move the instrument slightly to enable the robot to properly
place the material in the position.
"Contact Us" on page
Twister II Robot User's Manual
page
113.
7).
Caliper Life Sciences, Inc.
Troubleshooting 114