Cleaning The Rack Base; Changing The Twister Ii Fuse(S) - Caliper Life Sciences Twister II User Manual

Robot
Table of Contents

Advertisement

Cleaning the Rack Base

The surface beneath the Twister II racks must be clean at all times.
Debris or dust under the racks will mis-align the robot and the
racks.

Changing the Twister II Fuse(s)

Twister II Fuses:
NOTE
To order fuses, contact Caliper Technical Support (see
The Twister II can operate at 100/120V~ using a single 3Amp fuse
or at 240V~ using two 1.6A fuses. For instructions for the Assay
Engine Twister II, see
WARNING
To change the Twister II fuses or change the fuse configuration:
1
2
3
P/N 113564 Rev. 2
100/120V operation: 1 x F3A250V, 0.25 x 1.25" (P/N 36483)
240V operation: 2 x T1.6A250V, 5 x 20mm, (P/N 44232)
page
Electrical shock hazard. Disconnect the power cord before
changing the fuses.
For continued fire protection and correct functioning of the unit,
replace fuses only with exact part number to prevent fire.
Verify the Twister II power switch is OFF and the power cord is
unplugged.
Using a small flat-blade screwdriver, gently pry the cover/fuse
block assembly (see
housing.
Pry Here
Figure 83. Pry Open the Fuse Housing
Remove the fuse block/cover assembly from the housing.
Twister II Robot User's Manual
Twister II - Maintenance 105
108.
Figure
83) away from the power entry port
Caliper Life Sciences, Inc.
page
7).