Caliper Life Sciences Twister II User Manual page 112

Robot
Table of Contents

Advertisement

Issue 3
The Twister II does not properly place materials into an instrument or the
storage racks.
Possible Cause
The instrument was moved
The wrong rack is in a rack position
(e.g. Rack 3 is in rack position 2)
The plates are rubbing on the input
or output racks.
The Twister II is not lowering the
plates all the way down before
releasing them.
P/N 113564 Rev. 2
Solution
Re-teach the instrument positions (see
page
113).
Either place the proper rack in the position or
re-teach the storage positions (see
Re-teach the storage positions (see
Re-teach the storage positions (see
Make sure the storage configuration is correct
if using the Stack Storage resource for CLARA.
Twister II Robot User's Manual
Troubleshooting 112
page
113).
page
113).
page
113).
Caliper Life Sciences, Inc.