Changing the Assay Engine Twister II Fuses
Assay Engine Twister II Fuses:
•
NOTE
To order replacement fuses, please contact Caliper Technical
Support. See
The Assay Engine Twister II operates at 100 - 240V~ using two
2.5A fuses. For instructions for the Twister II, see
Twister II Fuse(s)" on page
WARNING
•
•
To change the Assay Engine Twister II fuses:
1
2
P/N 113564 Rev. 2
100 - 240V~ operation: 2 x T2.5A250V, 5 x 20mm (P/N 114703)
"Contact Us" on page
Electrical shock hazard. Disconnect the power cord before
changing the fuses.
For continued fire protection and correct functioning of the unit,
replace fuses only with exact part number to prevent fire.
Verify the Assay Engine Twister II power switch is OFF and the
power cord is unplugged.
Using a small flat-blade screwdriver at the top of the fuse
housing, gently pry the fuse housing (see
power entry port.
Pry Here
Figure 87. Assay Engine Twister II Power Entry Port
Twister II Robot User's Manual
Twister II - Maintenance 108
7.
105.
Figure
Caliper Life Sciences, Inc.
"Changing the