SENSe:SWEep Subsection

Sweep Time
[:SENSe]:SWEep:TIME <time>
[:SENSe]:SWEep:TIME?
Specifies the time in which the instrument sweeps the display.
Remarks:
A span value of 0 Hz causes the analyzer to enter zero
span mode. In zero span the X-axis represents time
rather than frequency. In this mode, the sweep time
may be set to faster values when Option AYX (fast
digitized time domain sweeps) is installed.
Range:
5 ms to 2000 seconds without Option AYX, fast
digitized time domain sweeps; 20 µs to 2000 seconds
with Option AYX, in zero span only
Default Unit:
seconds
Front Panel
Access:
Sweep, Sweep Time Auto Man
Automatic Sweep Time
[:SENSe]:SWEep:TIME:AUTO OFF|ON|0|1
[:SENSe]:SWEep:TIME:AUTO?
Automatically selects the fastest sweep time for the current settings.
Factory Preset
and *RST:
On
Front Panel
Access:
Sweep, Sweep Time Auto Man
Sweep Time Mode
[:SENSe]:SWEep:TIME:AUTO:MODE SRESponse|SANalyzer
[:SENSe]:SWEep:TIME:AUTO:MODE?
Specifies the type of automatic coupling for the fastest sweep time at
the current settings.
Stimulus response
Spectrum analyzer
Factory Preset
and *RST:
SANalyzer
Front Panel
Access:
Sweep, Sweep Coupling SR SA
Chapter 5
Language Reference
SENSe Subsystem
5-97