Measurement Functions - The Sense Subsystem; Hp 81635A And Hp 81619A- Master And Slave Channels - HP 8163A Series Programming Manual

Lightwave multimeter, lightwave measurement system & lightwave multichannel system, mainframes
Hide thumbs Also See for 8163A Series:
Table of Contents

Advertisement

Measurement Operations & Settings
command:
syntax:
description:
N O T E
parameters:
response:
example:
affects:
command:
syntax:
description:
parameters:
response:
example:
68
:SLOT[n]:HEAD[m]:TST?
:SLOT[n]:HEAD[m]:TST?
Returns the latest selftest results for an optical head.
This command does not perform a selftest. Use selfTeST command, "*TST?" on
page 49, to perform a selftest.
none
Returns an error code and a short description of the error.
slot:head:tst?
+0,"self test OK"<END>
Optical heads
:SPECial:REBoot
:SPECial:REBoot
Reboots the mainframe and all modules.
none
none
spec:reb
Measurement Functions – The
SENSe Subsystem
The SENSe subsystem lets you control measurement parameters for a Power
Sensor, an Optical Head Interface module, or a return loss module.
HP 81635A and HP 81619A- Master and
Slave Channels
For the HP 81635A Dual Power Sensor and HP 81619A Dual Optical Head
Interface module, channel 1 is the master channel and channel 2 is the slave
channel. The master and slave channels share the same software and hardware
triggering system. For some commands, setting parameters for the master channel
sets the parameters for the slave channel. In these cases, you may only set
parameters for the slave channel by setting master channel parameters.
The commands listed in Table 5 can only be configured using the master channel.
The commands listed in Table 6 are independent for both master and slave
channels.
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User's Guide, E1299
Measurement Functions – The SENSe Subsystem

Advertisement

Table of Contents
loading

This manual is also suitable for:

8164a series8166a series

Table of Contents