CONFigure Subsystem

CONFigure Subsystem
CONFigure subsystem commands apply only to measurements found in
the
menu. These commands stop the current measurement
MEASURE
and set up the instrument for the specified measurement using the
factory default instrument settings.
:CONF:<measurement> will always set :INIT:CONT OFF (single
mode), and also places the measurement in the idle state.
:CONF:SANalyzer causes the present measurement to exit
(functionally the same as pressing
analyzer in base instrument spectrum analyzer state.
The query :CONFigure? returns the current measurement name in
quotes.
Configure the Channel Power Measurements
:CONFigure:ACPower
Configure the Channel Power and Density Measurements
:CONFigure:CHPower
Configure the Emission Bandwidth Measurements
:CONFigure:EBWidth

Set Number of Measured Harmonics

:CONFigure:HARMonics[:<number>]
Set the number of harmonics to be measured.
Factory Preset
and *RST:
10
Range:
2 to 10, limited to the maximum frequency range of the
analyzer
Remarks:
This command sets the number of harmonics to
measure before computing the total harmonic
distortion. The minimum number is two (the
fundamental and the second harmonic).
Front Panel
Access:
MEASURE, Harmonic Dist, Meas Setup, Harmonics
Configure the OBW and Transmit Frequency Error
Measurements
:CONFigure:OBWidth
5-36
), and places the
MEASURE, Meas Off