HP ESA-E Series Programmer's Manual page 58

Esa spectrum analyzers
Table of Contents

Advertisement

HP ESA Status Registers
Using the Analyzer Status Registers
Figure 2-15
15
Bit Number
STATus:QUEStionable:CALibration:CONDition<num>
STATus:QUEStionable:CALibration:CONDition?
Questionable
2-28
14
13
12
11
10
9
Status
Calibration Condition Register
Bit
Description
0, 1
Reserved. These bits are not used by the analyzer, but are for
future use with other HP products.
2
A 1 in this bit position indicates that a failure has occurred while
trying to align the tracking generator (TG).
3
A 1 in this bit position indicates that a failure has occurred while
trying to align the RF section.
4
A 1 in this bit position indicates that a failure has occurred while
trying to align the IF.
5
A 1 in this bit position indicates that a failure has occurred while
trying to align the local oscillator (LO).
6
A 1 in this bit position indicates that a failure has occurred while
trying to align the analog-to-digital converter (ADC).
7
A 1 in this bit position indicates that a failure has occurred while
trying to align the FM demodulation circuitry.
8
A 1 in this bit position indicates that a failure has occurred while
trying to align the quasi-peak detector.
9
Unused. These bits are always set to 0.
10
A 1 in this bit position indicates that a tracking peak needs to be
performed for the CAL OUTPUT (the tracking generator is in
operation).
11
A 1 in this bit position indicates that the alignment of the RF
section was skipped, perhaps due to an external 50 MHz signal
having been detected.
8
6 5 4 3 2 1 0
7
ck752a