HP 8590-Series
Analyzers
Command(s)
BAUDRATE
:SYSTem:COMMunicate:SERial[1]|2|3|4|5|6|7|8
[:RECeive]:BAUD <baud_rate>
BAUDRATE?
:SYSTem:COMMunicate:SERial[1]|2|3|4|5|6|7|8
[:RECeive]:BAUD?
BIT
BITF
BLANK
BLANK TRA
:TRACe[1]|2|3:MODE BLANk
BLANK TRB
BLANK TRC
BML
:TRACe:MATH:SUBTract:DLINe <trace>
Corresponding SCPI Command(s)
Description/Comments
Specifies the baud rate of a spectrum analyzer
with the RS-232 interface option (Option 1AX)
installed.
Specifies the baud rate of the instrument, with
the RS-232 interface installed. If no optional
serial port number is specified, port 1 is
assumed. The Transmit baud rate is set to the
same value as the receive baud rate.
The HP 8590-Series analyzers return data in
the format: 1200. The HP ESA analyzers
return data in the format: +1200.
Returns the state of a bit.
Returns the state of a bit.
Blanks trace A, trace B, or trace C and stops
taking new data into the specified trace.
Selects the blank display mode for the selected
trace. TRACE1 corresponds to trace A,
TRACE2 corresponds to trace B, and TRACE3
corresponds to trace C. The blank display mode
turns off the trace data so that it is not viewed
on the display.
Subtracts display line from trace B and places
the result in trace B.
Subtracts DLINe (display line) from TRACE2
(trace B) and places the result into TRACE2
(trace B).