HP ESA-E Series Programmer's Manual page 176

Esa spectrum analyzers
Table of Contents

Advertisement

Language Reference
CALCulate Subsystem
Front Panel
Access:
Define Peak Search
:CALCulate:MARKer:PEAK:SEARch:MODE PARameter|MAXimum
:CALCulate:MARKer:PEAK:SEARch:MODE?
Sets the peak search mode.
See command :CALCulate:MARKer[1]|2|3|4:MAXimum
NOTE
Factory Preset
and *RST:
Remarks:
Front Panel
Access:
Define Peak Threshold
:CALCulate:MARKer:PEAK:THReshold <ampl>
:CALCulate:MARKer:PEAK:THReshold?
Specifies the minimum signal level for the analyzers internal peak
identification routine to recognize a signal as a peak. This applies to all
traces and all windows.
See command :CALCulate:MARKer:PEAK:EXCursion
NOTE
Range:
Default Unit:
Front Panel
Access:
5-22
Search, Search Param, Peak Excursn
MAXimum
If mode is set to MAXimum, peak search will place the
marker at the maximum amplitude in the trace. If
mode is set to PARameter, peak search will place the
marker at the highest peak that rises and falls by at
least the peak excursion above the peak threshold. If no
peak meets the excursion and threshold criteria, a No
Peak Found error (error 202) is issued.
Next peak, next peak right, next peak left, and peak
table are not affected by this command. They will
always use peak excursion and peak threshold for
search criteria.
Search, Search Param, Peak Search Max
Reference level to the bottom of the display
amplitude units
Search, Search Param, Peak Threshold
Chapter 5