HP ESA-E Series Programmer's Manual page 252

Esa spectrum analyzers
Table of Contents

Advertisement

Language Reference
SENSe Subsystem
Time Gating Delay
[:SENSe]:SWEep:TIME:GATE:DELay <time>
[:SENSe]:SWEep:TIME:GATE:DELay?
Sets the delay time from when the gate trigger occurs to when the gate
opens. This is for EDGE triggering only.
Factory Preset
1 µs
and *RST:
0.3 µs to 429 seconds
Range:
Default Unit:
seconds
Front Panel
Access:
Sweep, Gate Setup, Edge Setup, Gate Delay
Time Gate Length
[:SENSe]:SWEep:TIME:GATE:LENGth <time>
[:SENSe]:SWEep:TIME:GATE:LENGth?
Specifies the gate time length in seconds. For EDGE triggering only.
Factory Preset
1 µs
and *RST:
0.3 µs to 429 seconds
Range:
Default Unit:
seconds
Front Panel
Access:
Sweep, Gate Setup, Edge Setup, Gate Length
Time Gate Level
[:SENSe]:SWEep:TIME:GATE:LEVel HIGH|LOW
[:SENSe]:SWEep:TIME:GATE:LEVel?
Selects the level of the gate signal. This command is for LEVel
triggering only.
Factory Preset
and *RST:
High
Front Panel
Access:
Sweep, Gate Setup, Level Setup
Time Gate Polarity
[:SENSe]:SWEep:TIME:GATE:POLarity NEGative|POSitive
[:SENSe]:SWEep:TIME:GATE:POLarity?
Selects the polarity of the gate signal. This command is for EDGE
5-98
Chapter 5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents