Front Panel
Access:
Frequency Counter Marker
:CALCulate:MARKer[1]|2|3|4:FCOunt[:STATe] OFF|ON|0|1
:CALCulate:MARKer[1]|2|3|4:FCOunt[:STATe]?
:CALCulate:MARKer[1]:FCOunt:X?
Turns on or off the marker frequency counter. To query the frequency
counter, use :CALCulate:MARKer[1]:FCOunt:X?
Factory Preset
and *RST:
Remarks:
Front Panel
Access:
Marker Function
:CALCulate:MARKer[1]|2|3|4:FUNCtion BPOWer|NOISe|OFF
:CALCulate:MARKer[1]|2|3|4:FUNCtion?
Selects the marker function for the specified marker. To query the value
returned by the function, use :CALCulate:MARKer[1]|2|3|4:Y?
Remarks:
Front Panel
Access:
Marker Peak (Maximum) Search
:CALCulate:MARKer[1]|2|3|4:MAXimum
Performs a peak search based on the search mode settings of
:CALCulate:MARKer:PEAK:SEARch:MODE.
See command :CALCulate:MARKer:PEAK:SEARch:MODE
NOTE
Front Panel
Access:
Chapter 5
Freq Count, Resolution Auto Man
Off
If query with frequency count off, 9e15 is returned.
Freq Count, Marker Count On Off
BPOWer is the power integrated within the bandwidth
NOISe is a noise measurement
OFF turns off all functions
When a measurement under the front panel MEASURE
key is started, this command is turned off. If this
command is turned on when any of the MEASURE key
measurements are in progress, that measurement will
be stopped.
Marker, Function
Search, Meas Tools, Peak Search
Language Reference
CALCulate Subsystem
5-19