position the markers at the desired trace X values. The value is in the
X-axis units (which is often frequency or time). Use
:CALCulate:MARKer:MODE SPAN to select span markers.
The query returns the current X-value frequency span of the designated
markers.
Default Unit:
Matches the units of the trace on which the markers
are positioned
Front Panel
Access:
Marker, <active marker>, Span
Band-Markers Start Frequency X Value
:CALCulate:MARKer[1]|2|3|4:X:STARt <param>
:CALCulate:MARKer[1]|2|3|4:X:STARt?
Position the start (reference) frequency of the designated band-type
marker pair, at the specified trace X value. The value is in the X-axis
units (which is often frequency or time). Use
:CALCulate:MARKer:MODE BAND to select band markers.
The query returns the current X-value start/reference frequency of the
designated marker.
Default Unit:
Matches the units of the trace on which the markers
are positioned
Front Panel
Access:
Marker, <active marker>, Start
Band-Markers Stop Frequency X Value
:CALCulate:MARKer[1]|2|3|4:X:STOP <param>
:CALCulate:MARKer[1]|2|3|4:X:STOP?
Position the stop frequency of the designated band-type marker pair, at
the specified trace X value. The value is in the X-axis units (which is
often frequency or time). Use :CALCulate:MARKer:MODE BAND to
select band markers.
The query returns the current X-value stop frequency of the designated
marker.
Default Unit:
Matches the units of the trace on which the markers
are positioned
Front Panel
Access:
Marker, <active marker>, Stop
Chapter 5
Language Reference
CALCulate Subsystem
5-29