HP ESA-E Series Programmer's Manual page 328

Esa spectrum analyzers
Table of Contents

Advertisement

HP 8590-Series
Analyzers
Command(s)
MKFCR
MKFCR <freq>
:CALCulate:MARKer:FCOunt:RESolution <real>
MKFCR AUTO
:CALCulate:MARKer:FCOunt:RESolution:AUTO ON|1
MKFCR UP|DN
:CALCulate:MARKer:FCOunt:RESolution?
MKFCR?
MKMIN
:CALCulate:MARKer[1]|2|3|4:MINimum
Corresponding SCPI Command(s)
Description/Comments
Sets the resolution of the marker frequency
counter.
Sets the resolution of the marker frequency
counter. AUTO ON couples the marker counter
resolution to the frequency span.
Sets the resolution of the marker frequency
counter so it is automatically coupled to the
frequency span, generating the fastest
accurate count.
The HP 8590-Series analyzer outputs data in
the format: 1000. The HP ESA analyzer
outputs data in the format: +1000.
Moves active marker to minimum signal
detected.
Places the selected marker on the lowest point
on the trace that is assigned to that particular
marker number.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents