HP ESA-E Series Programmer's Manual page 46

Esa spectrum analyzers
Table of Contents

Advertisement

HP ESA Status Registers
Using the Analyzer Status Registers
Figure 2-3.
15
14
Bit Number
Operation
1
This refers to any measurement under the
The Operation Status Register continuously monitors the hardware
and firmware status of the analyzer, and is read-only. To query the
register, send the command :STATus:OPERation:CONDition? The
response will be the decimal sum of the bits which are set to 1. For
example, if bit number 9 and bit number 3 are set to 1, the decimal sum
of the 2 bits is 512 plus 8. So the decimal value 520 is returned.
The transition filter specifies which types of bit state changes in the
condition register will set corresponding bits in the event register. The
changes may be positive (from 0 to 1) or negative (from 1 to 0). Send the
2-16
13
12
11
10
9
8
STATus:OPERation:CONDition?
Status
Condition Register
Bit
Description
0
A 1 in this bit position indicates that the analyzer is
performing a self-calibration.
1,2
Unused. These bits are always set to 0.
3
A 1 in this bit position indicates that a sweep is in
progress.
4
A 1 in this bit position indicates that a measurement
in progress.
5
A 1 in this bit position indicates that a measurement
in a "wait for trigger" state.
6,7
Unused. These bits are always set to 0.
8–14
Reserved. These bits are not used by the analyzer, but are
for future use with other HP products.
15
Always Zero (0).
6 5 4 3 2 1 0
7
MEASURE