Error Messages
-199 to -100: Command Errors
-158
String data not allowed
A string data element was encountered, but not allowed
by the device at this point in the parsing.
-130
Suffix error
This error, as well as errors -131 through -139, are
generated when parsing a suffix. This particular error
message is used if the device cannot detect a more
specific error.
-138
Suffix not allowed
A suffix was encountered after a numeric element
which does not allow suffixes.
-134
Suffix too long
The suffix contained more than twelve characters (see
IEEE 488.2, 7.7.3.4).
-102
Syntax error
An unrecognized command or data type was
encountered. For example, a string was received when
the device does not accept strings.
-124
Too many digits
The mantissa of a decimal-numeric data element
contained more than 255 digits excluding leading zeros
(see IEEE 488.2, 7.7.2.4.1).
-113
Undefined header
The header is syntactically correct, but it is undefined
for this specific device. For example, *XYZ is not defined
for any device.