HP 8590-Series
Analyzers
Command(s)
SRCPSWP
:SOURce:POWer:SPAN <rel_ampl>
or,
:SOURce:POWer:SWEep <rel_ampl>
:SOURce:POWer:MODE FIXed|SWEep
SRCPSWP?
:SOURce:POWer:SWEep?
SRCPWR
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude] <ampl>
:SOURce:POWer:MODE FIXed|SWEep
SRCPWR?
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude]?
Corresponding SCPI Command(s)
Description/Comments
Selects sweep range of the source output.
Specifies the range of power levels through
which the source output will sweep. Use
:SOURce:POWer:STARt to set the power
level at the start of the power sweep.
Sets the source output to be at a single
amplitude (fixed) or to sweep through a
range of power levels.
The HP 8590-Series analyzer outputs data
in the format: 10.00. The HP ESA analyzer
outputs data in the format:
+1.00000000E+001.
Selects the source power level.
Specifies the source output power level and
turns on the source. An offset power level
can be added to the output power to
compensate for system losses between the
source and the receiver.
Sets the source output to be at a single
amplitude (fixed) or to sweep through a
range of power levels.
The HP 8590-Series analyzer outputs data
in the format: 10.00. The HP ESA analyzer
outputs data in the format:
+1.00000000E+001.