HP ESA-E Series Programmer's Manual page 314

Esa spectrum analyzers
Table of Contents

Advertisement

HP 8590-Series
Analyzers
Command(s)
GD
GD <value>
[:SENSe]:SWEep:TIME:GATE:DELay <time>
GD UP|DN
GD?
[:SENSe]:SWEep:TIME:GATE:DELay?
GDRVCLPAR
GDRVGDEL
GDRVGLEN
GDRVGT
GDRVGTIM
GDRVPRI
Corresponding SCPI Command(s)
Description/Comments
Sets the delay time before the gate opens.
Sets the delay time from when the gate trigger
occurs to when the gate opens. This is for
EDGE triggering only.
The HP 8590-Series analyzer outputs data in
the format: 1E-6. The HP ESA analyzer
outputs data in the format: +1.00000000E-006.
Clears the pulse parameters (pulse width,
pulse repetition interval, and reference edge)
for a time-gate measurement by setting the
pulse parameters to 0.
For the frequency window only, GDRVGDEL
sets the time delay from when the gate trigger
occurs to when the gate is opened.
Adjusts the gate length in both the time and
frequency windows.
Turns the gate on or off in the frequency
window.
Activates the gate trigger marker, and places it
at the given value.
Enters the specified value as the pulse
repetition interval.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents