To avoid electric shock when using the probe, please make sure that the insulated wire of
the probe is in good condition. Do not touch the metallic part of the probe when the
probe is connected to high voltage source.
4.3.8

Probe Compensation

When used for the first time, the oscilloscope probe must be compensated to match
the input characteristics of the oscilloscope channel to which it is connected. A non-
compensated or poorly compensated probes can introduce measurement inaccuracy
or errors. The compensation procedure is as follows:
1. Perform step 1, 2, 3, and 4 in
2. Check the displayed waveforms and compare them with waveforms shown in the
figure below.
3. Use the probe compensation adjustment tool provided in the accessories to adjust
the low-frequency compensation adjustment hole on the probe until the displayed
waveform is consistent with the "Perfectly compensated" waveform shown in the
figure above.
4.4

Product Overview

The following content provides a description of the front and rear panels as well as
the user interface of the HDO1000 series by taking 4-channel model HDO1104 as an
example.