Probe Compensation - Rigol MSO8000A Series Quick Manual

Digital oscilloscope
Hide thumbs Also See for MSO8000A Series:
Table of Contents

Advertisement

4. Set the probe attenuation to 10X, and then press
5. Observe the waveform on the display. In normal condition, the square waveform as
shown in
6. Use the same method to test the other channels. If the square waveforms actually
shown do not match that in the figure above, please perform "
Compensation
WARNING
To avoid electric shock when using the probe, please make sure that the insulated wire of
the probe is in good condition. Do not touch the metallic part of the probe when the
probe is connected to high voltage source.
7.7

Probe Compensation

When the probes are used for the first time, you should compensate the probes to
make them match the input channels of the oscilloscope. Non-compensated or
poorly compensated probes may cause measurement inaccuracy or errors. The probe
compensation procedures are as follows:
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Figure 7.7 To Use the Compensation Signal
Figure 7.8
should be displayed.
Figure 7.8 Square Waveform Signal
" introduced in the next section.
Compensation Signal Output Terminal
Ground Terminal
.
To Prepare for Use
Probe
15

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mso8074aMso8154aMso8204a

Table of Contents