5.7

Probe Compensation

When used for the first time, the oscilloscope probe must be compensated to match
the input characteristics of the oscilloscope channel to which it is connected. The
non-compensated or poorly compensated probe may cause measurement errors. The
compensation procedure is as follows:
1. Perform Step 1, 2, 3 and 4 in Function Inspection.
2. Check the displayed waveforms and compare them with the waveforms shown in
Figure 5.7
3. Use the probe compensation adjustment tool provided in the accessories to adjust
the low-frequency compensation adjustment hole on the probe until the displayed
waveform is consistent with the "Perfectly compensated" waveform shown in the
above figure.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
.
Over compensated
Figure 5.7 Probe Compensation
Perfectly compensated
Under compensated
To Prepare for Use
19