Preparing for Setup

3-5-1

Physical Inspection

Verify that the system arrived intact (visual inspection).
If the system has been damaged, please refer to "Damage in Transportation" on page
beginning of this manual.
Verify that all items are present (see the Packing List). Report any missing parts on the Post
Delivery Checklist.
3-5-2

EMI Protection

This unit has been designed to minimize the effects of Electro-Magnetic Interference (EMI). Many of the
covers, shields, and screws are provided primarily to protect the system from image artifacts caused by
this interference. For this reason, it is imperative that all covers and hardware are installed and secured
before the unit is put into operation.
See
2-2-3 "EMI Limitations" on page 2-7
Section 3-6

Completing the Setup

3-6-1

System Specifications

3-6-1-1
System Requirements Verification
Verify that the site meets the requirements listed in
Verify that the specifications below don't conflict with any on-site conditions.
3-6-1-2
Physical Dimensions
The physical dimensions of the Vivid 7 unit are summarized in
Table 3-8
Height
137.5 - 157.5
54.1 - 62
3-6-1-3
Mass (Weight) with Monitor and Peripherals
Table 3-9
Model
Vivid 7
3-6-1-4
Maximum Audio Noise
Less than 70dB(A) according to the standard DIN 45635 - 19 - 01 - KL2.
3 - 14
11
Physical Dimensions of Vivid 7 with monitor and peripherals
Width
Depth
64
25.2
Weight of Vivid 7 with monitor and peripherals
Weight [kg]
Weight [lbs]
200
Section 3-5 - Preparing for Setup
for more information about EMI protection.