System Software - Installation/Upgrade Procedure - GE Voluson E8 Service Manual

Hide thumbs Also See for Voluson E8:
Table of Contents

Advertisement

GE H
DRAFT
EALTHCARE
D
KTD102576, R
IRECTION
Section 8-2

System Software - Installation/Upgrade Procedure

8-2-1
Introduction
NOTICE
To update the Systems C:\ image via the FMI FROM DVD button in the System Setup
see:
Section 8-2-5 "System Software - Installation Procedure (FMI from DVD)" on page
!! NOTICE:
The Software parts to be upgraded (e.g., Ultrasound Application Software, Service Software, EUM,
MS Patches, etc.) depend on the contents of the System DVD being used.
NOTICE
If the currently installed software has to be upgraded to a newer version, and the system DVD being
used is SW 9.0.3 or higher, a new software specific "Permanent key is required.Please contact your
!! NOTICE:
local distributor or GE service representative.
8-2-2
Manpower
One person ~ 1 hour (depends on contents of System DVD, peripherals, etc.)
8-2-3
Tools
System DVD
8-2-4
Preparations
Before performing the Software Upgrade:
make sure that all system functions are working correct
check the current Application Software version
if the currently installed software has to be upgraded by a newer version, and the system DVD being
used is SW 9.0.3 or higher, calculate new software specific "Permanent key" in OKOS.
Please contact your local distributor or GE service representative to get the necessary key.
NOTE:
It is NOT necessary to save Full System Configuration (Full Backup) prior to the upgrade.
All existing User Programs, 3D/4D Programs and Auto Text settings remain untouched!