Main Console Parts Replacement; Fan Assembly Replacement - GE Vivid E80 Service Manual

Hide thumbs Also See for Vivid E80:
Table of Contents

Advertisement

Main Console parts replacement

Fan Assembly replacement

Read and follow
WARNING
Preparations
Remove the Fan Assembly
Vivid E80/E90/E95
GC091052 Rev. 3
Energy Control and Power Lockout for .
When servicing parts of the Ultrasound system where there is
exposure to voltage greater than 30 volts:
1. Follow LOCK OUT/TAG OUT procedures.
2. Turn off the breaker.
3. Unplug the Ultrasound system.
4. Maintain control of the Ultrasound system power plug.
5. Wait for at least 30 seconds for capacitors to discharge as
there are no test points to verify isolation.
Ultrasound System components may be energized.
When preparing for the replacement, you must perform the
following steps:
1. Power down the Ultrasound system.
2. Disconnect the Mains Power Cable from the wall outlet.
3. Disconnect all probes and external I/O cabling.
4. Remove the Right Side Cover.
Follow this procedure to remove the Fan Assembly:
1. Unscrew the screws of the Card Rack Cover.
2. Remove the Card Rack Cover.
3. Pull the Fan Assembly out of the frame.
Service Manual
Main Console parts replacement
8-69