Safety Issues; Probe/Connectors Check; Power Supply Adjustment - GE LOGIQ A3 Basic Service Manual

Table of Contents

Advertisement

GE H
EALTHCARE
D
5213326-100, R
IRECTION
Section 4-7Safety Issues
4-7-1

Probe/Connectors Check

Take the probes and check them as described below:
Table 4-11
Step
1.
Test Each delivered Probe
2.
Test each probe in each connector slot
Hold the probe connector vertically with the cable pointing
upward.
Turn the connector locking handle to the horizontal position.
Align the connector with the probe port and carefully push into
place.
3.
Rotate the locking handle to the full vertical position to lock in
place. Position the probe cable so that it is not resting on the floor
CAUTION:
Do not allow the probe head to hang freely. Impact to the probe
head may result in irrepairable damage.
Press the "Probe" key available on the control panel to activate
4.
second probe connected.
Rotate the lock handle counter-clockwise to the horizontal
position to unlock the connector.
5.
Remove the connector from the port
Ensure that the probe head is clean before placing the probe in its
storage case.
Probes can be connected at any time, whether the unit is on or off
WARNING
WARNING
Do NOT touch the patient and any of the connectors on the ultrasound unit simultaneously, including
ultrasound probe connectors.
WARNING
WARNING
Take the following precautions with the probe cables: Keep free from the wheels. Do
not bend.Do not cross cables between probes.
4-7-2

Power Supply Adjustment

There are no adjustments on the power supplies. The DC Power is self-regulated. If a voltage is outside
the specified range, it means that something is wrong, either with the power supply itself or with a
component connected to that specific power outlet.
Approv e d Docu m ent - 5213326-100TPH_r3.pdf Page 100 of 252
3
EVISION
Probe and connectors check
Task to do
it will display pictorial data on the screen
It will display pictorial data each time
Able to connect a probe
Second probe details loaded on the screen automatically.
To disconnect probes:
The probes that are not connected to the unit should be stored in
their storage case.
Chapter 4
LOGIQ™ A3 S
ERVICE
Expected Result(s)
M
ANUAL
4-13