General Procedures and Functional Checks

4-3-25 Probe/Connectors Check

NOTE:
4-3-25-1 To connect a Probe
CAUTION
CAUTION
Step
1
Select the appropriate connected probe from the
probe indicators on the Touch Panel.
2
Lauch the application. To change application,
press Probe key on the Control Panel.
3
Verify there's no EMI/RFI or artifacts specific to the
probe.
4-30
Probes can be connected at any time, whether the unit is ON or
OFF.
1. Place the probe's carrying case on a stable surface and
open the case.
2. Carefully remove the probe and unwrap the probe cable.
3. Put the porbe in the probe holder.
DO NOT allow the probe head to hang free. Impact to the
probe head could result in irreparable damage.
4. Hold the probe connector vertically with the cable pointing
upward.
5. Slide the connector lock to the left (unlocked position).
6. Align the connector with the probe port and carefully push
into place.
7. Slide the connector lock to the right position to secure the
probe connector.
8. Carefully position the probe cable in the probe cord holder
spot so it is free to move, but not resting on the floor.
TAKE THE FOLLOWING PRECAUTIONS WITH THE PROBE
CABELS:
-KEEP AWAY FROM THE WHEELS
-DO NOT BEND
-DO NOT CROSS CABLES BETWEEN PROBES
Table 4-1:
Probe and Connectors Checks
Task
Expected Results
The probe activates in the currently-selected
operating mode. The probe's default settings
for the mode and selected exam are used
automatically.
The selected application starts.
No EMI/RFI or artifacts.
Vivid T8/Vivid T8 Pro – Basic Service Manual
5718826-100 English Rev. 10