Operator's Service
Troubleshooting

To check the power-up tests

The logic analyzer automatically performs power-up tests when you
apply power to the instrument. The revision number of the operating
system shows in the upper-right corner of the screen during these
power-up tests. As each test completes, either "passed" or "failed"
prints on the screen in front of the name of each test.
1 Disconnect all inputs, then insert a formatted disk into the
flexible disk drive.
2 Let the analyzer warm up for a few minutes, then cycle power by
turning off then turning on the power switch.
If the analyzer is not warmed up, the power-up test screen will
complete before you can view the screen.
3 As the tests complete, check if they pass or fail.
The Flexible Disk Test reports No Disk if a disk is not in the disk drive.
Performing Power-Up Self-Tests
passed
passed
passed
passed
passed
passed
No Disk
480
ROM text
RAM test
Interrupt test
Display test
PS2 Controller Test
Hard Disk Test
Flexible Disk Test