System Performance Analysis (SPA) Software
System Performance Analysis Software
Zooming in on an area of interest. When viewing the State
Overview display, you may see areas of high activity and areas of little
or no activity. To zoom in on one of these areas for more resolution, put
the X and O markers on the boundaries of the area, then adjust the low
and high values to match the X and O marker positions.
Using State Histogram mode
Choosing a label to monitor. To specify a label to monitor, select
the Label field in the State Histogram display. In the pop-up, you will
see a list of all the labels defined in the Format specification. From this
list, choose the label you want to monitor.
NOTE:
Changing from one label to another will erase the display setup for the first
label. If you want to change to a different label, but do not want to lose the
setup for the current one, save the current one to a disk.
Defining the ranges. To define a range on the specified label, select
one of the 11 range fields. When you select one of the 11 ranges, you
will see the Range Definition pop-up. This pop-up has fields where you
enter the low and high value, a name for the range, and whether the
range is on or off.
SPA State Histogram Menu