Using the Trigger Menu
Arming and Additional Instruments
In this example STATE MACH triggers from Group Run, then arms
Example
TIME MACH and Scope.
To duplicate this, set STATE MACH to run from Group Run, TIME
MACH to run from STATE MACH, and Scope Arm In to Analyzer.
Arming with two analyzers and an oscilloscope
When the run starts, the state analyzer automatically begins evaluating
its trigger sequence instruction. When the trigger sequence is satisfied,
the state analyzer sends an "Arm" signal to the timing analyzer, the
oscilloscope, and the external trigger.