Operator's Service

Preparing For Use

Preparing For Use
This section gives you instructions for preparing the logic analyzer for
use.
Power requirements
The logic analyzer requires a power source of either 115 VAC or 230
VAC, -22 % to +10%, single phase, 48 to 66 Hz, 200 Watts maximum
power.
Operating environment
The operating environment is listed in previous chapter. Note the
noncondensing humidity limitation. Condensation within the
instrument can cause poor operation or malfunction. Provide
protection against internal condensation.
The logic analyzer will operate at all specifications within the
temperature and humidity range given. However, reliability is
enhanced when operating the logic analyzer within the following
ranges:
• Temperature: +20°C to +35 °C (+68 °F to +95 °F)
• Humidity: 20% to 80% noncondensing
Storage
Store or ship the logic analyzer in environments within the following
limits:
• Temperature: -40 °C to +75 °C
• Humidity: Up to 90% at 65 °C
• Altitude: Up to 15,300 meters (50,000 feet)
Protect the logic analyzer from temperature extremes which cause
condensation on the instrument.
473