User Mode
Basic Macros
Timing trigger macro library
The following list contains all the macros in the library of timing trigger
macros. They are listed in the same order as they appear onscreen.
User level - custom combinations, loops
The User level is a user-definable level. This level offers low-level
configuration and uses one internal sequence level. If the "<" duration
is used, four levels are required.
1. Find anystate "n" times
This macro becomes true when the first state it sees occurs "n" number
of times. It uses one internal sequence level.
2. Find pattern present/absent for > duration
This macro becomes true when it finds a designated pattern that has
been present or absent for greater than or equal to the set duration. It
uses one internal sequence level.
3. Find pattern present/absent for < duration
This macro becomes true when it finds a designated pattern that has
been present or absent for less than the set duration. It uses four or
five internal sequence levels.
4. Find edge
This macro becomes true when the designated edge is seen. It uses one
internal sequence level.
Logic Analyzer Reference
The Analyzer Trigger Menu
325