Logic Analyzer Reference

The Analyzer Configuration Menu

The Analyzer Configuration Menu

Type field

The Type field lets you configure the logic analyzer with either an
internal clock (Timing mode) or an external clock (State and SPA).
When the Type field is selected, the following choices are available.
Timing. When Timing is selected, the analyzer uses its own internal
clock to clock measurement data into the acquisition memory. This
clock is asynchronous to the signals in the target system. Fields relating
to external clocks such as the Analyzer Format menu's Master Clock do
not appear and certain menus such as Compare are not available.
The analyzer can only be configured with one timing analyzer. If two
are selected, the first will be turned off.
State. When State is selected, the analyzer uses a clock from the
system under test to clock measurement data into acquisition memory.
This clock is synchronous to the signals in the target system.
State Compare (1670E-series only). When State Compare is
selected, the Compare menu is available in the main menu selection.
For more details on Compare, see "The Compare Menu." State
Compare mode functions much like State mode, except that total
memory is reduced.
SPA. SPA stands for System Performance Analysis. It uses an external
clock like a state analyzer but measures overall system performance
rather than recording discrete activity. For more details on SPA, see
page 362.