The Analyzer Trigger Menu

The Analyzer Trigger Menu

Trigger sequence levels

Sequence levels are the definable stages of the total trigger
specification. Individual sequence levels are assigned using either a
predefined trigger macro or a user-level trigger macro. The total trigger
specification can contain both kinds of macro.
"Using the Trigger Menu" on page 74 for more on setting up a trigger.
See Also
Sequence level usage
Generally, you would think using one macro in one sequence level uses
up one of the available sequence levels. This may not always be the
case. Some of the more complex predefined macros require multiple
sequence levels. Keep this point in mind if you are near the limit on
remaining sequence levels. The exact number of internal levels
required per macro, and the remaining available levels, are shown
within the macro library list. User macros, however, use only one level
almost all the time. The only instance where multiple levels are used
with the User macro is when the "<" duration is assigned.