Preparing For Use - Agilent Technologies 16962A Service Manual

Logic analyzer
Table of Contents

Advertisement

Agilent 16962A Logic Analyzer
Service Guide
2

Preparing for Use

To inspect the module
18
To configure and install the module
18
To test the module
19
To clean the module
19
This chapter gives you instructions for preparing the logic
analyzer module for use.
Power Requirements
All power supplies required for operating the logic analyzer are
supplied through the backplane connector in the mainframe.
Operating Environment
The operating environment is listed on
page
15. Note the
non- condensing humidity limitation. Condensation within the
instrument can cause poor operation or malfunction. Provide
protection against internal condensation.
The logic analyzer module will operate at all specifications
within the temperature and humidity range given on
page
15. However, reliability is enhanced when operating the
module within the following ranges:
Temperature: +20C to +35C (+68F to +95F)
Humidity: 20% to 80% non- condensing
Storage
Store or ship the logic analyzer in environments within the
following limits:
• Temperature: - 40C to +75C (- 40F to +167F).
• Humidity: Up to 90% at 65C (+149F).
• Altitude: Up to 15,300 meters (50,000 feet).
Protect the module from temperature extremes which cause
condensation on the instrument.