Appendix C Maintenance
C

Configure the instrument settings

Configure the instrument settings

(Optional) Set
the instrument
name
86
After you install the QuantStudio
instrument settings for your region as explained below.
Note: The serial number of the QuantStudio
default instrument name. We recommend that you name the QuantStudio
Instrument uniquely. While not critical to network function, unique instrument names
can aid with sample tracking and instrument troubleshooting. The QuantStudio
Instrument does not test the uniqueness of the instrument name when it is set.
1. From the Main Menu of the touchscreen, touch
Settings.
2. In the Instrument Settings menu, touch Instrument Name.
3. In the Instrument Name screen, touch the Instrument Name field.
4. In the Edit Instrument Name screen, enter up to a 16-character name for the
instrument, then touch Enter.
IMPORTANT! The instrument name is an alphanumeric string used to identify
the QuantStudio
3D Instrument on the network. The instrument name:
·
Can be up to 16 characters long.
·
Must consist of alphabetic and numeric characters.
·
Can contain hyphens, but not as the first or last characters in the name.
·
Cannot include spaces or special characters ( ; : " < > * + = \ | ? , ).
5. Touch OK when done to save the setting, then touch
Menu.
3D Digital PCR Instrument, you can configure the
3D Digital PCR Instrument is the
, then touch Instrument
QuantStudio
3D Digital PCR System User Guide
3D
3D
to return to the Main