Introduction
1A.4.0 APPLICATIONS
RAD SYSTEMS:
Bucky table, table top and off-table radiography
Vertical bucky/cassette radiography
Conventional tomography
R&F SYSTEMS
Fluoroscopic and spot film applications
Tomography with conventional and/or remote R&F tables
Optional digital compatible.
1A.5.0 FEATURES
High frequency generator.
One or two tube operation, Rad or Rad / Fluoro.
Single or dual filament supplies.
Low speed or dual speed X-ray tube stator supply.
Optimal matching of X-ray tubes by PROMs.
Repetitive self checks of generator functions, provides display of system faults and operating errors.
Optional AEC, up to four inputs.
Optional ABS with kVp or kVp/mA fluoro stabilizer.
Optional remote fluoro control box for table top use or SFD mounting.
X-ray Tube protection. The generator allows setting the following limits:
a) Maximum mA, adjustable for each focal spot.
b) Maximum kVp, adjustable for each X-ray tube.
c) Maximum kW, adjustable for each focal spot.
d) Maximum filament current limit, adjustable for large and small focal spots.
e) Anode heat warning and anode heat alarm levels.
Calibration features :
a) Microprocessor design allows all calibration and programming to be performed via the Console.
b) mA calibration is automated.
Messages and diagnostic information: For users and service personnel, the generator console
displays various messages indicating status or equipment problems. The user is prompted in case of
errors.
Error log stores last 200 errors and associated generator settings.
Service and diagnostic information available via a lap-top computer (optional).
KVp range:
mA range:
mAs range:
Time range:
Refer to product description (end of section 1D) for compatibility and features of this specific generator.
Page 1A-4
Radiography 40 to 150 kVp.
Fluoroscopy 40 to 125 kVp.
Radiography 10 to 320 mA (30 kW), 10 to 400 mA (37.5 kW), 10 to 630 mA (50
kW), 10 TO 800 mA (65 kW) and 10 to 1000 mA (80 kW).
Fluoroscopy 0.5 to 6.0 mA.
tube dependent, max 1000 mAs.
Radiography 2.0 to 6300 ms.
Fluoroscopy 0 to 5 or 0 to 10 minutes.
Rev. E
Millenia / Indico 100 Series Service Manual
CPI Canada Inc.
Ch # 740904-00