5
Troubleshooting
5.2.3
Error Messages (Cont)
ERROR CODE
E058
E059
E060
E061
E062
E063
E070
E100
Page 5-8
MESSAGE
CONSOLE BUT ERR
HOUSE HEAT LIMIT
EXP KV HIGH
EXP KV LOW
EXP_ SW ERROR
FACTORY DEFAULTS
SOFTWARE KEY ERR
CAL_MAX MA ERR
Rev. B
Millenia / Indico 100 Series Service Manual
PROBLEM
Console has detected
exposure exceeded backup
time and terminated
exposure.
X-ray tube housing has
exceeded limit.
KV exceeds high KV
tolerance level.
KV exceeds low KV
tolerance level.
The EXP_SW signal on the
generator Interface and
generator CPU board is
enabled when it should be
disabled.
SW1 switch 8 on the
generator CPU board is set
to default the generator CPU
NVRAM with factory
defaults.
Defective or missing GAL
U29 on generator CPU
board 734573.
Maximum mA has been
exceed during auto
calibration.
CPI Canada Inc
ACTION
Call product support.
Wait for tube to cool.
1. Check the output
of the kV
reference DAC on
the generator
CPU board.
2. Measure the
output of the
generator with a
dynalyzer or a
non-invasive kVp
meter.
1. Check the output
of the kV
reference DAC on
the generator
CPU board.
2. Measure the
output of the
generator with a
dynalyzer or a
non-invasive kVp
meter.
Call product support.
Set SW1 switch 8 to
its non default
position. The
generator will not exit
the initialization
phase until this
switch is set.
Call product support
for new GAL U29.
Repeat auto
calibration and/or
decrease standby
current.
Ch # 740904-09