7.0 Aec Density Calibration - CPI Indico 100 Series Service And Installation Manual

Hide thumbs Also See for Indico 100 Series:

Advertisement

CPI Canada Inc.
3D.6.0 SHORT AEC TIME COMPENSATION (IF APPLICABLE) Cont
Step
7.
8.
9.
10.
11.
3D.7.0 AEC DENSITY CALIBRATION
FILM SCREEN 1
FILM SCREEN 2
FILM SCREEN 3
EXIT
*TUBE?
-8: -
-7: -
-6: -
-5: 62%
3200MS EXIT
*TUBE?
-4: 50%
-3: 37%
-2: 25%
-1: 12%
3200MS
<<
Millenia / Indico 100 Series Service Manual
Reduce the absorber thickness again such as to decrease the AEC exposure time to
approximately 6 ms (but not less).
Adjust the short AEC time compensation pot for the AEC channel being calibrated such
that the dose (or mAs) is approximately the same as it was in step 6.
The short AEC time compensation adjustments affect the AEC calibration at longer
exposure times. Therefore, it may now be necessary to readjust the gain pot for the AEC
channel being calibrated to restore the dose (or mAs) values to the values previously
recorded in table 3D-14. Ensure that the absorber thickness and mA values are as per
table 3D-13 when readjusting the AEC gain pot.
Films should be exposed and developed, and the O.D. checked at the 75 kV breakpoint
at AEC exposure times of approximately 6 ms and approximately 100 ms. If the film
density is not acceptable at both short and long AEC exposure times, it will be necessary
to iterate the adjustments of both the short AEC time compensation pot and the AEC
gain pot by repeating steps 3 to 8.
Repeat steps 1 to 10 for each image receptor (AEC channel) to be short AEC time
compensated.
The following screens are used for density calibration
* AEC CALIBRATION *
0%HU
*DENS. SE TUP*
0%HU
*DENS. SETUP*
Action
DENSITY SETUP
>>
>>
Ch # 740904-06
Rev. D
AEC Calibration
+
-
+
-
Page 3D-31
3D