3D.7.0 AEC DENSITY CALIBRATION (Cont)
Step
12.
13.
14.
15.
16.
Page 3D-34
Set the calibration number for the highest density plus step to the desired relative density
value (example 100, this will result in approximately double the 0 density dose).
Repeat step 8 and 9.
Select the next lowest density step (i.e. 7) and enter the appropriate calibration number
for that step. Refer to table 3D-15 and notes preceding table 3D-15. Repeat steps 8 and
9.
Repeat the previous step for each remaining density plus step. Use the >> and <<
buttons to navigate through the DENS SETUP menu in order to access all the density
steps.
When the density setup is complete and verified via dose measurements, use the <<
and EXIT buttons to return to the AEC CALIBRATION menu.
Rev. D
Millenia / Indico 100 Series Service Manual