Download Print this page

2.0 Compatible Devices And Options; 2.1 X-Ray Tubes; 2.2 Stators; 2.3 Aec Devices - CPI Indico 100 Series Service And Installation Manual

Hide thumbs Also See for Indico 100 Series:

Advertisement

1D
Compatibility Listing
1D.2.0 COMPATIBLE DEVICES AND OPTIONS
The Millenia and Indico 100 family of generators may be factory or user configured to be compatible with
various external devices. Certain features must be factory configured, others are user configurable.
Please refer to chapter 2 and 3 of this manual and/or consult the factory for further specifics.
1D.2.1 X-Ray Tubes
Various makes and models of inserts and housings are supported.
CAUTION:
PLEASE ENSURE THAT THE X-RAY TUBE INSERT IS AS STATED ON THE PRODUCT
DESCRIPTION.
IF THE HOUSING HAS BEEN RELOADED WITH ANOTHER INSERT TYPE, CALIBRATION
MAY BE SERIOUSLY AFFECTED AND MAY CAUSE TUBE DAMAGE.
1D.2.2 Stators
Various types/impedances of stators are supported.
1D.2.3 AEC Devices
Various AEC devices (ionization, solid state or PMT) may be supported via the optional AEC board,
depending on configuration.
1D.2.4 ABS Pickups
Various ABS pickups (light diode, composite video, PMT, etc) are supported on R&F generators
depending on configuration.
1D.2.5 Tomographic Tables
Various. Please note that the generator is used as a backup timer ONLY in tomography.
AEC is NOT available for tomography.
1D.2.6 Digital Interfaces (R&F Generators)
The generator may be configured to be compatible with various digital imaging systems
1D.2.7 Options
Major options include AEC board, remote fluoro control unit, dual speed starter, line adjusting transformer,
and two tube HT transformer.